|
MFE 1212 Mark II Operation and Maintenance Manual
Table of Contents
Section 1.
Introduction
Section 2.
Magnetic Flux Leakage
Section 3.
MFE 1212 Mark II General Description
Section 4.
MFE 1212 Mark II Operation
Section 5.
MFE 1212 Mark II Maintenance
Section 6.
Ultrasonic Prove Up
Section 7.
Battery Management.
This equipment makes use of very powerful rare
earth magnets. When disassembling the unit or moving the magnetic carriage great
care should be exercised to prevent damage to either the operator or equipment.
Maintenance work should always be carried out on a wooden as opposed to metal
bench. All tools should be kept well away from the magnetic bridge until
required.
SECTION 1
Introduction
This manual has been written to ensure that operators of the MFE 1212 Mark II have all the
information necessary to carry out the best possible examination of any given
tank floor. It is strongly recommended that any technician intending to use this
equipment reads this document in its entirety prior to carrying out any
inspections. Any operator using this equipment should
be capable of demonstrating a full understanding of the inspection principles
involved. Ultrasonic Prove Up should only be carried out by personnel who are
adequately trained, certified, and experienced in the evaluation of corrosion
type indications. More than one significant defect has been missed due to the
improper ultrasonic evaluation of indications detected by magnetic flux leakage.
The MFE 1212 Mk II is intended for use as a supplemental scanner to the
MFE 2412 Mk II. It uses the same handle and Electronics module as the MFE 2412
Mk II and these items can be interchanged to provide some redundancy. It is not
recommended that the 1212 be used in areas where it is possible to scan with the
2412 as the smaller bridge is not capable of achieving the same levels of
saturation and therefore sensitivity may be lower to smaller volume indications.
SECTION 2
Magnetic Flux Leakage
This equipment uses Magnetic Flux Leakage as a
detection tool. The powerful magnetic bridge introduces a magnetic flux into the
material near to saturation level. Any localized reduction in the thickness of
the material will result in a flux leakage at the surface. A series of sensors
are placed between the poles of the magnetic bridge to detect these leakage
fields. The strength of the leakage field is a function of volume loss and is
not a reliable indication of remaining wall thickness. Although the amplitude of
the signal generated by the sensors gives some relative severity information it
is not recommended that amplitude alone be used for accept/reject criteria
purposes. Flux Leakage should only be used as a detection tool. Truly
quantitative information can only be obtained using Ultrasonic assessment of the
areas identified by the Magnetic Flux Leakage Scanner.
SECTION 3
MFE 1212 Mark II General Description
The MFE 2412 Mark II is comprised of the following modules:-
1.
Magnet Bridge
2.
Sensor
3.
Battery Module
4.
Handle Assembly
5.
Electronics Module
6.
Signal and power cables
Magnetic
Bridge
The Magnetic Bridge provides the necessary flux
levels to achieve saturation of the plate thickness to be inspected. It is
supported on four low rolling resistance, durable wheels at each corner
allowing easy maneuverability of the scanning head. An additional set of guide
wheels are attached to one side of the bridge to maintain proximity to the shell
when inspecting the annular plates.
Sensor
Bar
The Sensor Bar contains an array of sensors
across the full width of the scanning head. These sensors detect the leakage
fields generated by the inspection process. The Sensor Bar extends almost to the
outer edge of the wheels allowing maximum possible coverage. It is suspended and
spring mounted within the Magnetic Bridge by two studs and adjustable wing nuts
allowing the adjustment of the height of the sensors from the inspection
surface.
Battery
Module
The Battery Module contains an 8 amp hour sealed
gel Lead Acid Battery that is both leak proof and can be placed in any
orientation. This type of battery is not regarded as hazardous by the airlines
for shipping purposes. The battery can provide power for the system for at least
8 hours of continuous use. A 3 amp fuse is located under the top cover to
prevent any damage to the module should a short occur within the power supply
wiring. The Battery Module can be quickly and easily
removed from the scanner for charging purposes.
Handle
Assembly
The Handle Assembly is connected to the Magnetic
Bridge using two 3/8 inch bolts, washers and nylock nuts and rests between two
sets of shoulder bolts. The handle provides mounting points in the form of
shoulder bolts and clamping bolts for the Battery Module and the Electronics
Module and can be folded in half for transportation. When in use it is locked in
place using two quick release pins. The leverage provided by the handle allows
the scanner to be easily tipped onto the rear wheels for removal or moving
around the floor.
Electronic
Module
The Electronic Module processes the signals from
the sensors and displays the data on a twelve channel L.E.D. display panel. Only
the first six channels are utilised in the MFE 1212 Mk II. The module is exactly
the same as that for the MFE 2412 Mk II and can be interchanged at will. The
front panel incorporates the controls for power, gain, alarm and display
brightness. A voltage display is incorporated to give a real time battery
voltage indication. If this drops below a nominal 12.0v reading then the battery
must be recharged using the battery charger provided. A fuse holder is mounted
on the back panel and contains a 3 amp fuse to protect the electronics in case
of a power problem.
SECTION 4
MFE 1212 Mark II Operation
Setup
The scanner is shipped in a custom
transportation case that contains the complete scanning system, function test
plate, battery charger, tool kit and spares. In order to unpack the container it
should be placed on it's back and the lid removed by releasing the catches and
using the handles provided. N.B.. The Function Test Plate is mounted in the lid
and can be removed by unscrewing the two retaining wing nuts and carefully
lifting it out. THE TWO HANDLES ATTACHED TO THE LID ARE TO ASSIST IN THE REMOVAL
OF THE LID ONLY AND ARE NOT TO BE USED TO LIFT OR CARRY THE COMPLETE
TRANSPORTATION CASE.
Remove the battery charger, tools and spares.
The complete handle assembly including the battery and electronic processor can
now be removed from the container. Extended the handle
to its full length and pin in place using the two quick release pins supplied.
Lift the container up on end with the magnetic bridge now on the bottom. At this
time the bridge assembly can be lifted from the keep plate by using the leverage
provided by the front handle grip.
Remove the the two attachment assemblies from
the pivot plate at the bottom of the handle. The 3/8" bolt, locknut washer and
spacer are used on the front connection point. The spacer is placed between the
bearing on the bridge and the top pivot plate. Tighten the bolt and locknut only
enough to take out any play in the connection.
The 3/8" clevis pin, washers and clip can be
inserted in the rear connection point to lock the scanner in the straight ahead
position.
Remove the function test plate from inside the lid.
Connect the cables to the Electronics Module
(these should be disconnected for shipping). They are keyed connectors and can
only be connected in the correct configuration. The small connector is for power
and the large connector for signal. Once everything is
connected, place the unit on the system function test
plate and check that the sensor bar is correctly located. This can be
accomplished by lowering the sensor bar, using the two wing nuts provided, until
the bar rests on the surface. Tension each wing nut and then add an additional
two turns on each to lift the sensor bar just above the surface of the plate.
Turn the unit on by activating the ON/OFF
switch. Activate the alarm by pressing the ALARM ON/OFF switch. The LED below
the Alarm will be illuminated. RESET the ALARM by pressing the ALARM RESET
switch. Check that the battery voltage reads at least 12.0v. If not, charge the
battery.
The unit is now ready for scanning the system
function test plate. Raise the gain control by using the low and high gain
switches on the front panel to 60 and push the scanner across the test plate at
a moderate walking pace. Three distinct responses should be noticed. All the
channels should respond to the notch in the plate and illuminate the full
display to maximum screen height. Both the 40% through wall simulated pit and
the 3/16" through hole should illuminate at least one channel to full screen
height.
Check that both indications trigger the alarm at
this gain setting. Both red LED's above the display should illuminate and flash
if any signal goes into the red section of the display. The alarm can be
canceled by pressing the ALARM RESET switch. The alarm can be disabled by
pressing the ALARM ON/OFF switch and the amber LED will be extinguished.
If
the above function test is successful then the system is ready for scanning.
If this system check is carried out in the
actual tank, the test plate must be elevated at least 4" from the floor plate
using sufficient wooden supports to prevent the plate from bending.
Scanning
The scanner has been optimized for a medium
walking pace. Faster scanning speeds will result in higher signal amplitudes due
to the sensor characteristics and is not recommended and may be detrimental to
the inspection process. Slower speeds, however will result in lower signal
amplitudes, and it is recommended that when it is found necessary to scan at
lower speeds, e.g.. in confined spaces, the system gain must be increased to
compensate.
The scanner can be used with the pivot locked
for straight scanning or unlocked when scanning around the radius of the shell.
The guide wheels can be switched over to the other side of the bridge if
clockwise scanning is preferred.
Maximum coverage of the floor area is achieved
using overlapping scans on a plate by plate basis. The scanner is extremely
maneuverable and can be rotated around its axis and easily lifted from the plate
onto the rear wheels for quick and easy relocation purposes. Do not attempt to
pull or push the unit over the lap welds without first tipping the unit back
onto the rear wheels. Damage to the Sensor Bar and Magnetic Bridge can result
from this practice.
Any given floor will generate a level of noise
dependent on the top surface condition of the plate. This will be displayed as a
general low level signal across the display. The level of this noise will
dictate the achievable sensitivity of the inspection.
As soon as discreet signals are noted they must
be proved up using Ultrasonics. This will give the operator the information
required as far as the system sensitivity is concerned and allow the
determination of an adequate scanning sensitivity for the remainder of the
floor. Only the first six channels of the display are active. The system can be
rocked backwards and forwards over the indication for accurate location
purposes. Lines drawn at the center of the array and the channel number of the
indication will intersect within an inch of the cause of the indication.
All indications determined to be significant
by the operator must be marked up for subsequent prove up. When it is apparent
from the display that a large area of lake type corrosion is present, the
boundaries of such an area can be quickly determined using the scanner and the
most significant indications within that area marked up for prove up. Amplitude
alone must not be used solely as the criteria for prove up purposes. The
operator must use all the information available from the display to determine
the relative significance of any given indication.
It is therefore recommended that the provided
alarm facility is NOT used as the basis of accept/reject decisions but rather as
an alarm for the operator in the event that his attention is distracted from the
display whilst scanning.
SECTION 5
MFE 1212 Mark II Maintenance
**** CAUTION ****
This equipment
incorporates proprietary information and trade secrets and is the subject of a
"Restricted use, Nondisclosure and Non circumvention agreement signed by each
purchaser of this equipment. In order not to breach this agreement please take
note of the following restrictions.
The scanner has been designed to be as
maintenance free as possible and is extremely ruggedly constructed. The
inspection environment is often extremely dirty and dusty and therefore the bulk
of the maintenance required involves cleaning and removal of magnetic debris and
product residue. DO NOT SPRAY WATER OR DETERGENT ON ANY PART OF THE SCANNER.
DO NOT USE ANY SOLVENT CLEANERS ON THE DISPLAY TOUCH PANEL.
The
scanner can easily be dismantled with the use of the hand tools provided and the
individual parts cleaned using degreased and absorbent cloths. DO NOT REMOVE THE
STAINLESS STEEL COVER PLATES ON THE BRIDGE POLE PIECES. This will invalidate the
warranty. DO NOT ATTEMPT TO OPEN THE ELECTRONICS MODULE. This will invalidate
the warranty. DO NOT ATTEMPT TO DISMANTLE THE SENSOR BAR other than for the
removal and replacement of the sensor protector. This will invalidate the
warranty.
Should any of these items require attention they must
be returned to the manufacturer for service.
IT IS RECOMMENDED THAT ALL MAINTENANCE AND
CLEANING OPERATIONS CARRIED OUT ON THE BRIDGE BE CONDUCTED ON A WOODEN OR NON
MAGNETIC SURFACE FOR SAFETY REASONS
A.
As required during
scanning
1.
Remove build up of ferritic debris from the underside of the scanner.
2.
Check condition of Bridge and Sensor protection plates.
B.
Daily
1.
Run system function test.
2.
Check tightness of all fasteners.
3.
Check condition of wheels. (Any unevenness will result in vibration and
increased noise levels)
C.
Weekly
1.
Separate the handle from the Magnetic Bridge and remove the wheels to aid
in a thorough cleaning of the component parts of the
scanner. The Battery and Electronic Modules can be quickly released from the
handle for separate cleaning purposes.
2. Remove Sensor Protector cover, check for damage and replace in
necessary. Clean and dry the sensor bar.
The
wheel bolt threads and sensor protector retaining screws should be lubricated
with an anti-sieze compound periodically to ease removal.
SECTION 6
Ultrasonic Prove Up
Magnetic Flux Leakage is an excellent detector
of under floor corrosion but can give no truly accurate assessment of remaining
wall thicknesses due to the limitations imposed by the inspection environment.
Accurate quantitative assessment is obtained using Ultrasonic prove up.
It is important that the operator who carries
out this assessment is adequately trained and experienced in this type of
indication analysis. This is not just " thickness measurement ". In order to
arrive at accurate remaining wall thickness it is necessary for the operator to
use equipment which is capable of giving him the information that he needs as
well as the application of the correct technique. The following type of prove up
equipment and technique is strongly recommended:-
A.
It is essential that a good
quality A-scan display is used for the assessment of any corrosion
detected . The equipment must be capable of determining location, size and the
accurate quantification of remaining wall thickness.
B.
A good quality 5 Megahertz
twin crystal focused contact transducer of either 0.375" or 0.500" diameter.
C.
Surprisingly enough ordinary water and plenty of it
provides the best coupling media in most cases.
D.
The amplitude of the signal from the corroded back wall must be raised to
at least the same screen height as the calibration reflector.
E.
Elevated gates tied to a digital read out are not
recommended as this technique invariably
underestimates the severity of the indication.
F.
The nearest facet of the
reflection must be identified and the true remaining wall thickness read from
the timebase.
SECTION 7
Battery Management
The batteries provided with the units are sealed
gel lead acid and provide sufficient power for at least 8 hours of continuous
use even at very low temperatures when fully charged. They should only be
charged using the supplied battery chargers which are fully automatic and will
maintain the battery at full charge if left connected and turned on for extended
periods without damage to either battery or charger. If the charger is turned
off for any reason the battery
must be disconnected from the charger.
The fully automatic maintenance charger provided
with the scanner is the only charger that should be used on this equipment.
Charging of the battery can be achieved by plugging the connector plug from the
charger into the mating receptacle on top of the Battery Module.
Under no circumstances may the charger be rigged
so that the unit is powered up during the charging process. This will cause
severe damage to the electronics module.
It is recommended that the battery be charged
overnight, every night when the unit is in continuous use and continuously when
not in use. This will ensure an adequate maintenance of battery power so that
the unit will be ready for use at a moments notice.
If it is necessary to remove or replace the
battery at any time be sure to replace all of the rubber shielding and foam
provided to ensure the original degree of protection against physical damage and
possible dead shorts of the electrical system is maintained.
|